Standard: ASTM C518-04, Test Method for Determining Steady State Heat Flux and Heat Transfer Characteristics by Heat Flow Meter Method; GB10295-88 Determination of steady-state thermal resistance and related characteristics of insulation materials (heat flux meter method).
Technical Parameter
1. Thermal conductivity range: Plate sample 0.001-5W/mk, accuracy ± 3%,
2. Hot surface temperature: Room temperature -99.99 ℃, measured using a high degree digital display meter, with an accuracy of 0.2 level and a resolution of 0.01 ℃.
3. Cold surface temperature: 5-99.99 ℃, measured using a high degree digital display meter, with a precision of 0.2 level and a resolution of 0.01 ℃.
4. Hot surface temperature control: room temperature -99.99 ℃, using a high degree programmable digital display meter for temperature measurement, with a precision of 0.2 level and a resolution of 0.01 ℃. It is equipped with a flat plate heater and bidirectional thyristor control.
5. The cold surface is controlled by a high-precision constant temperature water tank.
6. Using computer automated testing.
7. Heat flow rate range: 0.5-2000W/m2, with a braiding rate of 0.25 W/m2.
8. Heat transfer area of hot and cold plates: 400 × 400 (mm).
9. Adjustable spacing between hot and cold plates: 0-100 (mm).
Instrument completeness
1. One main unit of DRPL-400 thermal conductivity tester;
2. 1 set of testing software (including communication interface);
3. 1 set of constant temperature water tank and its controller;
4. 1 branded desktop computer (may be provided by the purchaser)