Features
◆ Secondary biosafety cabinet, air flow mode: 30% discharge,
70% internal circulation.
◆ It conforms to nsf49 and en12469 standards, and the
product executive standard is the pharmaceutical industry standard of the people's
Republic of China, YY 0569-2011.
◆ The outer box body is made of high-quality cold-rolled
steel plate and electrostatic spray plastic; the three sides of the operation
area are made of stainless steel integrated structure, and the inner parts
that can be cleaned are made of 10 mm large fillet for easy cleaning.
◆ Two ULPA ultra-high efficiency filters are equipped as
standard, which can achieve 99.999% closure efficiency for 0.12 μ m
particle system. The filter membrane is made of non barrier borosilicate
glass fiber.
◆ The control software system is independently developed and
has obtained the national software copyright certificate. The LCD screen can
display the system parameters such as the working area temperature, air flow
rate, service life of the filter membrane, and can adjust the parameter
settings. The blue screen digital indicates the parameter values, and the
password modification program settings to prevent misoperation.
◆ Unique negative pressure leak proof design, the work area
surrounded by negative pressure, truly achieve zero leakage, to ensure
the cleanliness of the operation area.
◆ The fan is equipped with voltage fluctuation compensation
function and resistance induction compensation function, effectively
extending
the service life of ultra-high efficiency filter.
◆ A high-sensitivity and high-precision breeze speed sensor
is equipped at the working area and the external air outlet to detect the
wind speed
in real time.
◆ Temperature sensor: it can detect and display the
temperature in real time.
◆ Equipped with filter pre failure, damage and blocking
alarm technology, equipped with fan overload alarm prompt.
◆ Equipped with height limit alarm device, when the mobile
glass door exceeds the specified height limit (200mm), the user will be alerted
by sound and light.
◆ It is equipped with a working window opening limit alarm
system. When the opening height of the front window of the safety cabinet exceeds
the set height, when the flow velocity of the down airflow and the flow
velocity of the inflow airflow of the safety cabinet fluctuates by more
than 20% of its nominal value, the interlocking system will start the alarm
prompt.
◆ With UV appointment timing function.
◆ The front window integrated sliding door can be positioned arbitrarily,
and can be completely closed for sterilization. The front window glass
door shall be made of safety toughened glass with a thickness of no less than
6mm.
◆ Interlock design of front window and germicidal lamp: when
the front window is opened, the germicidal lamp is off.
◆ It has the function of parameter memory and call recovery
protected by power failure and data loss of dead state.
◆ The front suction inlet adopts the design of unimpeded
return air to ensure that the formation of counter current is effectively prevented
and the safety is improved during the operation.
◆ 10 degree tilt design, more in line with ergonomic power,
making the operator more comfortable.
◆ Power supply in the cabinet: waterproof socket design, the
whole machine has power-off protection function, making the operation more
flexible and convenient.
◆ The manufacturer has relevant qualification and patent
◆ (1)The company has passed ISO9001 quality management system
certification.
◆ (2)The manufacturer has obtained the medical device production
license.
◆ (3)The product has obtained SFDA medical
device registration certificate.
◆ (4)The control software obtained the
national software copyright certificate.
◆ (5)The product has dozens of patents.
◆ Before leaving the factory, the equipment shall be tested
by a qualified third party to ensure that the equipment is qualified to
delivery.
Specifications
Parameter / model |
BSC-1000ⅡA2 |
BSC-1300ⅡA2 |
Cleanliness grade |
HEPA: ISO level 5 (Class 100 of 100) /
ULPA: ISO level 4 (class 10 of 10) |
HEPA: ISO level 5
(Class 100 of 100) / ULPA: ISO level 4 (class 10 of 10) |
Filter level / efficiency |
HEPA:≥99.995%,@0.3μm/ULPA:≥99.999%,@0.12μm |
HEPA:≥99.995%,
@0.3μm/ULPA:≥99.999%, @0.12μm |
Descent wind speed |
0.33m/s |
0.33m/s |
Suction air speed |
0.53m/s |
0.53m/s |
Noise |
≤65dB |
≤65dB |
Vibration half peak |
≤5μm |
≤5μm |
Power / maximum power consumption |
AC 220V/50HZ / 1KW |
AC 220V/50HZ / 1KW |
Biosafety personnel protection |
CFU of all impact samplers ≤ 10, CFU of
slit samplers ≤ 5 |
CFU of all impact
samplers ≤ 10, CFU of slit samplers ≤ 5 |
Product protection |
CFU of all sampling dishes ≤ 5 |
CFU of all sampling
dishes ≤ 5 |
Cross contamination protection |
CFU of all culture dishes ≤ 2 |
CFU of all culture
dishes ≤ 2 |
Workspace size |
1000*520*640 mm |
1300*520*640 mm |
Overall dimension of device |
1165*760*2200 mm |
1465*760*2200 mm |
Specification of air supply filter |
990*470*70*① |
1290*470*70*① |
Specification of exhaust air filter |
650*450*90*① |
830*450*90*① |
Specification and quantity of
fluorescent lamp / UV lamp |
30W*②/30W*① |
40W*②/40W*① |
Illuminance |
≥900LX |
≥900LX |
Weight |
240kg |
270kg |
Applicable number |
1 |
1-2 |
Wind direction |
Top out |
Top out |
Parameter / model |
BSC-1600ⅡA2 |
Cleanliness grade |
HEPA: ISO level 5
(Class 100 of 100) / ULPA: ISO level 4 (class 10 of 10) |
Filter level / efficiency |
HEPA:≥99.995%,@0.3μm/ULPA:≥99.999%,@0.12μm |
Descent wind speed |
0.33m/s |
Suction air speed |
0.53m/s |
Noise |
≤65dB |
Vibration half peak |
≤5μm |
Power / maximum power consumption |
AC 220V/50HZ / 1KW |
Biosafety personnel protection |
CFU of all impact
samplers ≤ 10, CFU of slit samplers ≤ 5 |
Product protection |
CFU of all sampling
dishes ≤ 5 |
Cross contamination protection |
CFU of all culture
dishes ≤ 2 |
Workspace size |
1600*520*640 mm |
Overall dimension of device |
1765*760*2200 mm |
Specification of air supply filter |
1590*470*70*① |
Specification of exhaust air filter |
1130*450*90*① |
Specification and quantity of
fluorescent lamp / UV lamp |
40W*②/40W*① |
Illuminance |
≥900LX |
Weight |
300kg |
Applicable number |
2-3 |
Wind direction |
Top out |