Standards:GB/T3810.8-2006 Determination of linear thermal expansion of ceramic tiles, GB/T16535-2008 Test method for linear thermal expansion coefficient of fine ceramics: Top rod method, GB/T16920-2015 Determination of average linear thermal expansion coefficient of glass, GB/T3074 (1). 4-2003 Determination of thermal expansion coefficient of graphite electrodes, GB/T 7320-2008 Test method for thermal expansion of refractory products. National standard for pharmaceutical packaging materials YBB002003-2015- Determination of average linear thermal expansion coefficient
Main technical parameters
1. Adopting vertical and push rod type, easy to load and unload samples;
2. Maximum operating temperature: 1000 ℃, with an adjustable heating rate of 0-50 degrees per minute;
3. Voltage: AC 220V; Heating furnace power: 2000W;
4. Strong compatibility with samples, capable of measuring the expansion coefficient of glass rods, tubes, plates, and finished products. Sample size: Ф (6-32) × (30-70) mm (glass bottle body within 30ml) or Ф (5-81) × H (110-182) mm (500ml inner glass bottle), other sizes can be customized according to customer requirements; The oversized sample can be easily disassembled and cut into the above shape; (Ampoules, penicillin bottles, and other glass bottle bodies can be directly used as samples for determination without the need for cutting)
5. The instrument can be displayed online with a computer and can print out the expansion curve, all raw data, and Tf points, and calculate the linear expansion coefficient and Tg point;
6. Temperature fluctuation in the heating furnace: ≤ ± 1 ℃;
8. Temperature control accuracy: from room temperature to 600 ℃, the temperature control accuracy is ± 1 ℃; 600 ℃ to 1000 ℃, with a temperature control accuracy of ± 2 ℃;
9. The measurement resolution is 0.1 micrometers, and the testing accuracy is ± a × zero point one × 10-6K-1;
10. The software can automatically calculate the linear expansion coefficient and obtain the expansion rate curve, Tg, and Tf points when processing data.
11. Test frame material: quartz.
12. Heating element: high-temperature resistance wire.
Configurations
One vertical push rod dilatometer testing host, one set of testing software, one quartz standard sample, one air compressor, one standard sample, and one desktop computer (configured according to the contract).