Standards
GB/T17650.1-1998
GB/T17650.2-8, IEC60754-1:1994, IEC60754-2:1991
GB/T19666-2005 GA306.1-2007 Article 6.3.4, GA306.2-2007 article 6.3.4, UL2556-2007 Article 9.10 and 9.11. T
Technical specifications
1.Tubular furnace
Effective length of heating chamber and pipe diameter 600×¢42mm, equipped with an adjustable electrical heating system. Heating power: 220V/10KW
2. Combustion tube
Material: silica made of refractory tube, corrosion-resistant gas.
Specifications: the inner diameter of 40mm, the length of the tube furnace inlet side is 60 ~ 200mm, the length of the outlet is 60 ~ 100mm,leave a gap for heating expansion.
3. Combustion boat
Material: fused quartz tube
Specifications: length 100mm, width 25mm, depth 10mm
The combustion boat feeding mode: the use of bar magnet and platinum wire pulling mode
4. The collection of gas
The selection of two bottles; One was fitted with a magnetic stirrer and contained 220mL of sodium hydroxide melt.
5. Gas supply system
Air flow meter: 20mL ~ 200mL (adjustable)
Compressor (Provide for yourself)
6. Sampling
500mg~1000mg
7. Determination of halogen acid content: optional
Analytical balance: accuracy is 0.1mg(customer supplied)
Digital PH meter: accuracy is 0.02PH
Conductivity meter: Measuring range: (0 ~ 1×105) uS/cm