Standard: Guidelines for Drug Stability Testing in the 2020 Pharmacopoeia and Relevant Articles of GB/T10586-2006 Manufacturing
Technical Parameter
|
Drug Stability Test Chamber |
Comprehensive drug stability test box |
Drug strong light stability test box |
||
Model |
LHH-80SD LHH-150SD LHH-250SD |
LHH-80SDP LHH-150SDP LHH-250SDP |
LHH-150GSD LHH-150GSP LHH-250GSD LHH-250GSP |
LHH-150GP LHH-250GP LHH-400GP |
|
Temperature Control Range |
0~65℃ |
0~65℃(Illuminated15~50℃) |
|||
Temperature fluctuation/uniformity |
±0.5℃/±2℃ |
||||
Humidity range/deviation |
25~95%RH/±3%RH |
NO |
|||
Light intensity/error |
NO |
0~6000LX adjustable ≤ ± 500LX (infinite dimming) |
|||
Time range |
1-99 hours per segment |
||||
Temperature and humidity control methods |
BTHC |
Balanced temperature regulation method |
|||
Refrigeration system/refrigeration method |
Two sets of independent original imported fully enclosed compressors automatically switch (LHH-80SD/80SDP is a set of independent original fully enclosed compressors) |
||||
Controller |
Programmable touch screen controller |
||||
Sensor |
Pt100 platinum resistor Capacitive humidity sensor |
Pt100 platinum resistor |
|||
Ambient Temperature |
+5~30℃ |
||||
Power supply |
AC220V±10% 50HZ |
||||
Operating power |
About 650W/680W/750W |
About 680W/680W/750W/750W |
About 450W/450W/600W |
||
power |
2000W/2800W/3000W |
2950W/2950W/3200W/3200W |
1450W/1700W/3200W |
||
Volume |
80L 150L 250L |
80L 150L 250L |
150L 250L |
150L 250L 400L |
|
Load tray (standard) |
2/3/3 pieces |
2/3/3 pieces |
3/3 pieces |
3/3/4 pieces |