Standard:
GB/T 5990-2006 "Test method for thermal conductivity of shaped insulating refractory products (hot wire method)";
GB/T 17106-1997 "Test method for thermal conductivity of refractory materials (parallel line method)"
Main technical indicators
1. Measurement range: DRX-3A (cross line method) 0.015~2.0 W/(m · K)
DRX-3A (parallel line method) 0.015~5.0 W/(m · K)
DRX-3B (parallel line method) 0.015-25.0 W/(m · K)
2. Temperature range: Conventional testing is at room temperature, and different constant temperature devices and testing hotlines can be selected according to user requirements for testing at different temperatures. The cost will be calculated separately, as follows:
-40-120 ℃ (requires high and low temperature constant temperature box) with copper constantan hot wire
l Room temperature~1000 ℃ (requires distribution of resistance wire furnace) with nickel chromium nickel silicon or platinum platinum rhodium hot wire
l Room temperature~1300 ℃ (requires silicon carbide rod resistance furnace) with platinum platinum rhodium hot wire
l Room temperature~1600 ℃ (requires silicon molybdenum rod resistance furnace) with platinum platinum rhodium hot wire
3. Measurement accuracy:<± 5.0%
4. Sample size: 2 pieces not less than 40mm × 80mm × A cylinder composed of 114mm symmetrical rectangular prism or two semi cylindrical blocks with a cross-sectional diameter of not less than 80mm and a length of not less than 114mm stacked together; (or customized according to user requirements),
5. Temperature resolution: 0.001 ℃
6. Hotline power supply: DRX-3A current resolution 0.01mA, voltage resolution 0.1mV
DRX-3B current resolution 0.1mA, voltage resolution 0.1mV
7. Equipment power supply: voltage 220V; Frequency 50Hz; Power 500W (refer to the instruction manual of the electric furnace temperature controller for the requirements of using the electric furnace temperature controller power supply),
Standard configuration
including one DRX-3 testing host, one software, one communication interface and data cable set, one desktop computer, and one room temperature testing hotline set. Additionally, a constant temperature device and high-temperature wire heating are optional and configured according to the contract.