Standard:
GB/T 5990-2006 "Test method for thermal conductivity of shaped insulating refractory products (hot wire method)";
GB/T 17106-1997 "Test method for thermal conductivity of refractory materials (parallel line method)";
Main technical indicators
1. Measurement range: 0.015~25.0 W/(m · K)
2. Temperature range: -40~120 ℃ (requires high and low temperature constant temperature box) with copper constantan hot wire
Room temperature~1000 ℃ (requires distribution of resistance wire furnace) with nickel chromium nickel silicon or platinum platinum rhodium hot wire
Room temperature~1300 ℃ (requires silicon carbide rod resistance furnace) with platinum platinum rhodium hot wire
Room temperature~1600 ℃ (requires silicon molybdenum rod resistance furnace) with platinum platinum rhodium hot wire
3. Measurement accuracy:<± 5.0%
4. Sample size: 2 pieces not less than 40mm × 80mm × A cylinder composed of 114mm symmetrical rectangular prism or two semi cylindrical blocks with a cross-sectional diameter of not less than 80mm and a length of not less than 114mm stacked together; (or customized according to user requirements),
5. Temperature resolution: 0.001 ℃
6. Hotline power supply: current resolution 0.1mA, voltage resolution 0.1mV
7. Equipment power supply: voltage 220V; Frequency 50Hz; Power 500W (refer to the instruction manual of the electric furnace temperature controller for the requirements of using the electric furnace temperature controller power supply)
Standard configuration:
including one DRX-3B testing host, one software, one set of communication interface and data cable, one desktop computer, and one set of parallel hotlines for room temperature testing. Additionally, a constant temperature device and high-temperature wire heating are optional and configured according to the contract.