Model: MJ-MC01
Features
◎ The instrument has high sensitivity, wide measuring range and good reproducibility.
◎ Few chemical reagents are used; Less appliances and devices are used; Do not use inert gas.
◎ The specially made bubble bottle is very simple, and there is no mercury loss during operation.
◎ The unique circuit design makes the instrument have high stability and sensitivity, and the baseline almost has no drift.
◎ The measuring range is wide, and the samples with concentration of 0.01~100 μ g/L can be continuously measured under the same state.
◎ It is easy to operate, the computer prompts the whole operation process, and the intelligent data processing greatly reduces the workload of the experimenters.
Specifications
Measuring range |
0~100ug/g |
Sample mass (mg) |
0.1 |
Detection method |
Cold atomic absorption method |
Detection limit |
≤0.05μg/L |
Linear error |
±10 |
Relevance |
R>0.999(0.01~10 ug/L) |
Reproducibility |
Measured mercury standard solution, 0.04-100 ug/L, standard deviation<3% |
Stability |
The mercury standard solution prepared by the same reagent can meet the requirement of measurement error without retesting the calibration curve. |
Repeatability |
≤3 |
Baseline stability |
±0.002/3 min (at A = 0) |
Power supply |
AC220V±10% ,50Hz |
Power waste |
22W |
Net weight |
8.5Kg |
Overall dimensions |
450×350×160 (mm) |